Skip to main content

Trilogy 100 Portable Ventilator

Combines unique flexibility with ease of use and advanced technology

Trilogy 100 Portable Ventilator

Model No.: 1032800

Trilogy 100 Portable Ventilator

Versatile, easy-to-use Trilogy 100 is designed for home, hospital and acute-care used to provide invasive and non-invasive ventilator support for adult and pediatric patients.

A pressure and volume mode ventilator with proprietary algorithms, Trilogy 100 meets the changing needs of patients by combining unique flexibility with ease of use and advanced technology. Trilogy provides ventilation data and prescription management through the use of an SD memory card.

Since its launch in 2009, Trilogy has demonstrated clinical success and offered freedom and mobility to over 200,000 patients worldwide.

 

Product Details:

Model Number 1032800
Dimensions 16.68cm (L) x 28.45cm (W) x 23.52cm (H)
Weight 5kg
Package includes 1 x Trilogy 100 Ventilator
Manufacturer Philips Respironics
Additional Information

Battery: Internal Battery with 3 hour external battery combined capability of 11 hours

Controls:

  • Circuit types: Active with PAP, passive
  • Ventilation types: Volume control, pressure control, bilevel
  • Volume modes: AC, SIMV (w/PS), CV
  • Pressure modes: CPAP, S, S/T, T, PC-SIMV (w/PS)
  • IPAP: 4 – 50 cm H20
  • EPAP: 0 – 25 cm H20 for active circuits; 4 – 25 cm H20 for passive circuits
  • CPAP: 4 – 20 cm H20 for passive circuits
  • PEEP: 0 – 25 cm H20 for active circuits; 4 – 25 cm H20 for passive circuits
  • Pressure support differential: 0 – 30 cm H20

Measured Patient Parameters:

  • Tidal volume: 50 – 2000 ml
  • Breath rate: 0 – 60 for AC mode, 1 – 60 for all other modes
  • Inspiratory time: 0.3 – 5.0 secs
  • Rise time: 1 – 6
  • Ramp start pressure: 0 – 25 cm H20 for active circuits; 4 – 25 cm H20 for passive circuits; 4 – 19 cm H20 in CPAP mode
  • Ramp length: Off, 5 – 45 minutes
  • Flex: Off, 1 – 3
  • Flow trigger sensitivity: 1 – 9 l/min
  • Flow cycle: 10% – 90%
  • Apnea rate: 4 – 60 BPM
  • Tidal volume: 0 – 2000 ml
  • Minute ventilation: 0 – 99 l/min
  • Leak rate: 0 – 200 l/min
  • Respiratory rate: 0 – 80 BPM
  • Peak inspiratory flow: 0 – 200 l/min
  • Peak inspiratory pressure: 0 – 99 cm H20
  • Mean airway pressure: 0 – 99 cm H20
  • Percentage patient triggered breaths: 0 – 100%
  • I:E ratio: 9.9:1 to 1:9.9

Alarms:

  • Circuit disconnect: Off, 10 – 60 secs
  • Apnea: Off, 10 – 60 secs
  • High tidal volume: Off, 50 – 2000 ml
  • Low tidal volume: Off, 50 – 2000 ml
  • High minute ventilation: Off, .1 – 99 l/min
  • Low minute ventilation: Off, .1 – 99 l/min
  • High respiratory rate: Off, 4 – 80 BPM
  • Low respiratory rate: Off, 4 – 80 BPM